Medical Device Directive 93 42 EEC

ANNEX III – EC TYPE-EXAMINATION

Information & Training. | Medical Devices   >>>

 

1. EC type-examination is the procedure whereby a notified body ascertains and certifies that a representative sample of the production covered fulfils the relevant provisions of this Directive.

2. The application includes:

 
the name and address of the manufacturer and the name and address of the authorized representative if the application is lodged by the representative,
the documentation described in Section 3 needed to assess the conformity of the representative sample of the production in question, hereinafter referred to as the ‘type’, with the requirements of this Directive. The applicant must make a ‘type’ available to the notified body. The notified body may request other samples as necessary,
a written declaration that no application has been lodged with any other notified body for the same type.
 

3. The documentation must allow an understanding of the design, the manufacture and the performances of the product and must contain the following items in particular:

 
a general description of the type, including any variants planned, and its intended use(s),
design drawings, methods of manufacture envisaged, in particular as regards sterilisation, and diagrams of components, sub-assemblies, circuits, etc.,
the descriptions and explanations necessary to understand the above-mentioned drawings and diagrams and the operation of the product,
a list of the standards referred to in Article 5, applied in full or in part, and descriptions of the solutions adopted to meet the essential requirements if the standards referred to in Article 5 have not been applied in full,
the results of the design calculations, risk analysis, investigations, technical tests, etc. carried out,
a statement indicating whether or not the device incorporates, as an integral part, a substance, or human blood derivative, referred to in Section 7.4 of Annex I, and the data on the tests conducted in this connection which are required to assess the safety, quality and usefulness of that substance, or human blood derivative, taking account of the intended purpose of the device,
a statement indicating whether or not the device is manufactured utilizing tissues of animal origin as referred to in Directive 2003/32/EC,
the solutions adopted as referred to in Annex I, Chapter I, Section 2,
the pre-clinical evaluation,
the clinical evaluation referred to in Annex X,
the draft label and, where appropriate, instructions for use.
 

4. The notified body must:

4.1. examine and assess the documentation and verify that the type has been manufactured in conformity with that documentation; it must also record the items designed in conformity with the applicable provisions of the standards referred to in Article 5, as well as the items not designed on the basis of the relevant provisions of the abovementioned standards;

4.2. carry out or arrange for the appropriate inspections and the tests necessary to verify whether the solutions adopted by the manufacturer meet the essential requirements of this Directive if the standards referred to in Article 5 have not been applied; if the device is to be connected to other device(s) in order to operate as intended, proof must be provided that it conforms to the essential requirements when connected to any such device (s) having the characteristics specified by the manufacturer;

4.3. carry out or arrange for the appropriate inspections and the tests necessary to verify whether, if the manufacturer has chosen to apply the relevant standards, these have actually been applied;

4.4. agree with the applicant on the place where the necessary inspections and tests will be carried out.
5. If the type conforms to the provisions of this Directive, the notified body issues the applicant with an EC type-examination certificate. The certificate must contain the name and address of the manufacturer, the conclusions of the inspection, the conditions of validity and the data needed for identification of the type approved. The relevant parts of the documentation must be annexed to the certificate and a copy kept by the notified body.

In the case of devices referred to in Annex I, Section 7.4, second paragraph, the notified body shall, as regards the aspects referred to in that section, consult one of the authorities designated by the Member States in accordance with Directive 2001/83/EC or the EMEA before taking a decision. The opinion of the competent national authority or the EMEA must be drawn up within 210 days after receipt of valid documentation. The scientific opinion of the competent national authority or the EMEA must be included in the documentation concerning the device. The notified body will give due consideration to the views expressed in this consultation when making its decision. It will convey its final decision to the competent body concerned.

In the case of devices referred to in Annex I, Section 7.4, third paragraph, the scientific opinion of the EMEA must be included in the documentation concerning the device. The opinion of the EMEA must be drawn up within 210 days after receipt of valid documentation. The notified body will give due consideration to the opinion of the EMEA when making its decision. The notified body may not deliver the certificate if the EMEA’s scientific opinion is unfavorable. It will convey its final decision to the EMEA.

In the case of devices manufactured utilizing tissues of animal origin as referred to in Directive 2003/32/EC, the notified body must follow the procedures referred to in that Directive.

 

6. The applicant must inform the notified body which issued the EC type examination certificate of any significant change made to the approved product.

Changes to the approved product must receive further approval from the notified body which issued the EC type-examination certificate wherever the changes may affect conformity with the essential requirements or with the conditions prescribed for use of the product. This new approval must, where appropriate, take the form of a supplement to the initial EC type examination certificate.

 
  1.  Administrative provisions
7.2. Other notified bodies may obtain a copy of the EC type-examination certificates and/or the supplements thereto. The Annexes to the certificates must be made available to other notified bodies on reasoned application, after the manufacturer has been informed.

7.3. The manufacturer or his authorized representative must keep with the technical documentation copies of EC type-examination certificates and their additions for a period ending at least five years after the last device has been manufactured. In the case of implantable devices, the period shall be at least 15 years after the last product has been manufactured.

 

Information & Training.

Medical Device:

      • Validation. Classification. Regulation. Requirements. Current best practices.
      • FDA cGMP’s. EU MDR’s / MDD’s.
      • FDA Medical Device Regulation. Outline of the FDA regulatory requirements.
      • FDA Medical Device Classification. The FDA approach to Medical Device Classification.
      • EU Medical Device Regulation and Classification (per MDD’s).
      • New European Medical Device Regulations (MDR’s). MDR Classification. MDR General Safety requirements.
      • Current Good Manufacturing Practices. QSR’s. General requirements of the QSR’s.
      • Quality System requirements to maintain compliant Validations.
      • Medical Device Process Validation. Validation requirements. Protocol development. IQ. OQ. PQ.
      • Medical Device Software Validation.
      • Medical Device Design Validation.
      • Electronic Signature, Electronic Records.
      • Life Cycle Approach to Validation.
      • Risk Identification. Documentation. DHR’s. DMR’s.
      • Etc. Etc. …
      • Information & Training presentation >>>